Application Range:
- Standard Measuring Range: 2 to 60 ppm
- Number of Strokes n: 20
- Time for Measurement: Approx. 8 min
- Standard Deviation: ± 10 to 15 %
- Color Change: White → Brown Grey
Ambient Operating Conditions:
- Temperature: 0 to 40 °C
- Absolute Humidity: 1 to 15 mg H2O / L
Reaction Principle:
- C6H6 + I2O5 + H2SO4 → I2
Cross Sensitivity:
- Alkyl benzenes such as toluene or xylene up to a concentration of 200 ppm do not affect the indication.
- It is impossible to measure benzene in the presence of petroleum hydrocarbons and carbon monoxide