Draeger hydrogen Sulphide 0.2 %/a (0.02 to 7 vol% ) Short Term Measuring Tubes ( Pack of 10 )

0.68

SKU: Ch28101 Category:

Description

Application Range:

  • Standard Measuring Range: 0.2 to 7 vol. %
  • Number of Strokes n: 1 + 2 desorption strokes in clean air
  • Time for Measurement: Approx. 2 min
  • Standard Deviation: ± 5 to 10 %
  • Color Change: Pale Blue → Black

Ambient Operating Conditions:

  • Temperature: 0 to 60 °C
  • Absolute Humidity: max. 40 mg H2O / L

Reaction Principle:

  • H2S + Cu2+ → CuS + 2 H+

Cross Sensitivity:

  • In the presence of sulfur dioxide, the indicating layer can change to a yellowish color, but the hydrogen sulfide measurement is not affected.
  • Comparable concentrations of mercaptan will interfere with the reading.